Cell and battery analyzer ACK2.5.10.8. Current from 2.5uA to 2.5A (up to 20A). Voltage from -4.5 to +4.5V. 8 channels.

Analyzers of batteries АСК2.5.10.8 are intended for testing electrochemical cells, accumulators of any type, supercapacitors, fuel cells, redox cells and any other chemical current sources.

Характеристики
Voltage on tested cell
from -4.5 to +4.5 V
Charge and discharge current
from 2.5 uA to 2.5 A
Number of channels
8
Current for all channels
20 A

Description

Cell and battery analyzer ACK2.5.10.8 tests various sources and energy storage devices, for example:

Cell and battery analyzer ACK2.5.10.8 contain a controlled power supply and also a controlled electronic load. The power supply can operate in voltage, current or power stabilization modes. The load can operate in voltage, current or power stabilization modes, and also in resistance stabilization mode.

Due to this, cell analyzer ACK2.5.10.8 can perform many tests. For example, the simplest, such as a single charge or a single discharge of the cell. In addition, the instrument can perform cyclic tests, with multiple alternation of charge and discharge. Finally, electrochemical cell analyzer ACK2.5.10.8 can perform complex combination modes that simulate real-life cell conditions.

Each channel of the ACK2.5.10.8 analyzer has a very wide operating range of the charge and discharge current of the cell under test. From 2.5 micro Amperes to 2.5 Amperes. Thanks to this, the potentiostat ACK2.5.10.8 allows testing both factory batteries and laboratory electrochemical cells.

Battery analyzer ACK2.5.10.8 operates under computer control. During the test, the analyzer measures many parameters of the cell under test. The instrument returns measurement results in the form of text result files. In addition, the device software automatically generates charge-discharge graphs for the cell under test. Finally, the analyzer software automatically generates graphs of degradation of cell parameters (which is convenient for cycle-life tests).

Key parameters of cell and battery analyzer ACK2.5.10.8

Number of independent channels: 8;

Charge current as well as discharge current of the cell under test on each channel: from 2.5 μA to 2.5 A;

Charge current and discharge current of the tested cell with parallel operation of the channels: from 20 μA to 20 A;

Operating voltage range on the tested cells: from minus 4.5 to plus 4.5 V.

Detailed information about analyzer of cells АСК2.5.10.8